Compaq Graphical User Interface mar345 Manuel d'utilisateur Page 39

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 40
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 38
m
ar345 Program Manual 3 9
_
____________________________________________________________________________________
_
A
ppendix
A
. Configuration File
Each scanner has a scanner specific configuration file. This file must reside in directo
ry
$MARTABLEDIR and is called config.XXX, where XXX is the three−digit serial numbe
r.
Please refrain from changing values unless there is good reason to do so. The configuratio
n
file is a keyworded Ascii−file. Please refer to the man page mar345_config_file forafu
ll
description of keywords.
A typical configuration file looks like this:
! ______________________________________
!
! Configuration file for scanner no. 120
! ______________________________________
!
! Only modify the following keywords, if desired...
!
DIST MIN 75.30 MAX 426.30
!
INTENSITY MIN 2.0 WARN 20.0
!
USE SPY
USE STATS
USE INCR
USE RUN
USE HTML
USE SUMMARY
!
COLORS 64
SETS 4
!
WAVE 1.541789
MONOCHROMATOR Mirrors
GENERATOR Rotating anode
!
NETWORK PORT 4441 HOST 192.0.2.1
!
MODE 2300 ROFF 180 ADC 50 AADD −42 BADD −45
MODE 2000 ROFF 180 ADC 50 AADD −42 BADD −45
MODE 1600 ROFF 180 ADC 50 AADD −42 BADD −45
MODE 1200 ROFF 180 ADC 50 AADD −42 BADD −45
!
MODE 3450 ROFF 180 ADC 50 AADD −42 BADD −45
MODE 3000 ROFF 180 ADC 50 AADD −42 BADD −45
MODE 2400 ROFF 180 ADC 50 AADD −42 BADD −45
MODE 1800 ROFF 180 ADC 50 AADD −42 BADD −45
!
!IP−Diameter: 345mm 300mm 240mm 180mm
!GAPS 89615 78601 63852 49103
! 02.2.2000: GAPS valid for S/N 120
!
Vue de la page 38
1 2 ... 34 35 36 37 38 39 40

Commentaires sur ces manuels

Pas de commentaire