Compaq MC-1 Manuel d'utilisateur Page 79

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 110
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 78
Messages and Diagnostic Codes Messages and Diagnostic Codes
Release 0 34-ST-25-20 MC Toolkit User Manual 71
7/03
Messages and Diagnostic Codes
Messages and Diagnostic Codes
Table 24 MC Toolkit Error Messages
MC TOOLKIT ERROR MESSAGES
MESSAGE CORRECTIVE ACTION
Another Secondary master has been
detected! HART Protocol does not support
more than one secondary master on the
HART Network.
Remove one of the secondary masters from the network.
Checksum error on Modem Response! A noisy environment can cause this error. Repeat the command
again.
Com Port Read Timeout! Ensure that the Honeywell supplied modem cable is used and that
the connections are secure.
Make sure the modem has a good battery.
Com read Error!
Error writing to Com Port!
Error writing to Com Port!
EscapeCom function Error!
GetCommState Error!
Make sure ActiveSync is not running.
If several programs are active, try closing one or more open
programs.
Stop the MC Toolkit application by doing File | Exit and restart the
program.
Use the Pocket PC hardware reset. See your Pocket PC
documentation reset the unit. Caution: If the Symbol PPT 2800
computer is reset, it will lose all its installed programs and data. The
program will have to be installed again.
Make sure you are using a Honeywell approved Pocket PC.
Host Failed to get good response from the
Modem!
Ensure that the Honeywell supplied modem cable is used and that
the connections are secure.
Make sure the modem has a good battery.
Invalid Command number from the
Modem!
Communications problem between the Pocket PC and the modem.
Repeat the command again.
Invalid Handle Value! Make sure ActiveSync is not running.
If Several programs are active, try closing one or more open
programs.
Stop the MC Toolkit application by doing File | Exit and restart the
program.
Use the Pocket PC hardware reset. See your Pocket PC
documentation reset the unit. Caution: If the Symbol PPT 2800
computer is reset, it will lose all its installed programs and data. The
program will have to be installed again.
Make sure you are using a Honeywell approved Pocket PC.